Olympus BX53M Instructions Manual page 36

System microscope
Table of Contents

Advertisement

a
30
Restoring the stored brightness (LIM function)
to the factory default setting
1
Set the main switch of the microscope frame to
2
If the control box is combined, remove it.
3
Set the main switch of the microscope frame to (ON) while pressing
both the LIM switch a and the SET switch b . Keep pressing the
b
LIM switch a and the SET switch b . After 5 seconds, the stored
brightness returns to the factory default setting.
TIP
4
Set the main switch of the microscope frame to
control box is combined, attach it again.
For procedures to attach the control box, see "Attaching the control
box" (page 100).
If you release the LIM switch and the SET switch within
5 seconds, the stored brightness (LIM function) does not
return to the factory default setting and the microscope starts
normally.
(OFF).
(OFF). If the

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents