Setting The Movement Limit With The Coarse Focusing Knob; Method To Find The Approximate Focal Point Easily - Olympus BX53M Instructions Manual

System microscope
Table of Contents

Advertisement

a
4

Setting the movement limit with the coarse focusing knob

This function prevents the collision between the sample and the
objective and also simplifies the focusing.
After the sample is brought into focus by the coarse focusing knob,
if you turn the pre-focusing lever a in the arrow direction to lock
it, the upper limit of the coarse focusing knob is set at the locked
position.
In case you focus again after replacing the sample, the sample can
be brought into focus approximately by raising the coarse focusing
knob to the stop position. Then, adjust finely using the fine focusing
knob.
Note, the stage movement is not locked by the fine focusing knob.
If the pre-focusing lever is locked, the stage is not lowered
NOTE
to the lower limit.
5

Method to find the approximate focal point easily

a
When the sample thickness is not known
1
Turn the coarse focusing knob to lower the stage to the lower limit.
2
Move the stage so that the center of the stage comes to the front of
the scale a .
3
Place the sample on the center of the stage.
4
Turn the coarse focusing knob to raise the stage so that the top surface
of the sample comes to 0 mm of the scale a when looking at the
sample horizontally.
5
Move the desired area of the sample under the objective.
When the sample thickness is known
1
Place the sample on the stage.
2
Turn the coarse focusing knob to move the stage so that the top surface
of the stage comes to the "Sample thickness" position of the scale a .
If the height adapter (BX3M-ARMAD) is combined, the
NOTE
scale is not available.