Olympus BX53M Instructions Manual page 85

System microscope
Table of Contents

Advertisement

Cause
The setting of the dip switch is not correct.
The motorized nosepiece did not reach the click position
within the specified time.
The motorized nosepiece is damaged.
The control box (BX3M-CB) is not connected.
The control box (BX3M-CB) is damaged.
The hand switch (BX3M-HS) is not connected.
The hand switch (BX3M-HS) is damaged.
The MIX slider for reflected light observation (U-MIXR) is not
connected.
The MIX slider for reflected light observation (U-MIXR) is
damaged.
The cable for U-MIXR (U-MIXRCBL) or the extension cable
for U-MIXR (U-MIXRECBL) is disconnected.
Remedy
Set the main switch to
(OFF), and use the dip switch on the
side of the control box (BX3M-CB) to set the number of objective
mounting holes on the nosepiece.
Set the main switch to
(OFF) and check that there is no
obstacle when the nosepiece rotates.
If the nosepiece contacts with the obstacle, remove the obstacle.
Then, set the main switch to (ON). If the same phenomena still
occurs, contact Olympus.
Set the main switch to
(OFF) and connect the control box and
the AC adapter of the control box again.
Then, set the main switch to (ON) and confirm that the beep is
heard once.
If the indicator on the back of the controller does not light even
though the control box and the AC adapter of the control box
are connected again, and also if the beep sound is not heard
when the main switch is set to (ON), contact Olympus.
Set the main switch to
(OFF) and connect the hand switch
(BX3M-HS). If the indicator3 on the back of the control box does
not light even though the hand switch is connected, contact
Olympus.
If the indicator3 on the back of the control box does not light
even though the hand switch (BX3M-HS) is connected, contact
Olympus.
Connect the MIX slider for reflected light observation.
If the indicator2 on the back of the control box does not light
even though the MIX slider for reflected light observation is
connected, contact Olympus.
If the indicator2 on the back of the control box does not light
even though the MIX slider for reflected light observation is
connected, contact Olympus.
Contact Olympus.
BX53M
Page
16, 100
16
16, 100
, 117
16, 100
, 117
16, 115
115
105
105
79