Olympus BX53M Instructions Manual page 103

System microscope
Table of Contents

Advertisement

b
c
2
4
3
e
d
This picture shows the procedure to attach the
right hand control 150 mm x 100 mm stage
(U-SIC64). The same procedure also applies when
attaching other stages.
2
Attaching the stage
a
NOTE
1
1
2
3
4
NOTE
NOTE
The movable area of the right hand control 150 mm x 100
mm stage (U-SIC64) is fixed by factory default. Perform
followings before attaching the stage.
· Remove the tape on the side of the stage a .
· Move the Y-axis lock lever b in the arrow direction to
unlock the lever.
Turn the coarse focusing knob to lower the stage holder c sufficiently.
Loosen the stage clamping screw.
Align the stage positioning pin e with the groove d on the front of
the stage holder and fit the stage from above.
Tighten the stage clamping screw to secure the stage.
In order to prevent the stage knob from interfering the
knobs of the focusing section, following cautions are
required depending on the stage to be combined.
4x4 inch stage
(U-SIC4R2/U-SIC4L2)
Stage with coaxial knobs
on the bottom right (left)
(U-SVRM/U-SVLM)
When replacing the stage, remove the objective together
with the nosepiece in advance.
BX53M
Attach the stage so that the
knob comes to the front.
Attach the fine focusing knob
of the focusing section at the
opposite positions of the stage
knob.