Olympus BX53M Instructions Manual page 113

System microscope
Table of Contents

Advertisement

b
a
5
shows the procedure to attach the
This picture
reflected LED lamp housing (BX3M-LEDR).
The same procedure applies when attaching the
mercury lamp housing.
Attaching the reflected LED lamp housing or the mercury lamp housing
(during darkfield observation) to BX3M-URAS-S
1
2
NOTE
2
3
4
5
NOTE
6
7
Loosen the mounting screws a (2 positions) of the DF converter
(U-RCV) using the Allen screwdriver.
Be careful, if the mounting screw is loosened too much, it
may be come off.
Insert the light source for reflected light illumination into the light source
mounting hole of the DF converter until it touches the end.
Tighten the mounting screws a (2 positions) of the DF converter
using the Allen screwdriver.
Loosen the mounting screws b (2 positions) of the reflected light
illuminator using the Allen screwdriver.
Insert the light source for reflected light illumination attached with the
DF converter into the light source mounting hole of the reflected light
illuminator until it touches the end.
· When attaching the reflected LED lamp housing (BX3M-
LEDR), attach it so that the cable for the reflected LED lamp
housing comes to the right side when looking from the
back of the microscope frame.
· When attaching the mercury lamp housing, attach it so that
the radiation fin faces up.
Tighten the mounting screws b (2 positions) of the reflected light
illuminator using the Allen screwdriver.
If the reflected LED lamp housing is attached, connect the cable to
the connector on the back of the microscope frame. For details, see
"Cable connections" (page 115).
If the mercury lamp housing is attached, connect the cable to the
power supply (U-RFL-T).
For the cable connection of the mercury lamp housing, refer to
the instruction manual provided with the unit.
BX53M
107