Stage - Olympus BX53M Instructions Manual

System microscope
Table of Contents

Advertisement

Cause
The time for pressing the LIM switch and the SET switch on
the front of the microscope frame is short. (5 sec. or less)

5-5 Stage

Phenomena
a) When the stage is touched by
the hand, the image moves
significantly.
b) The stage does not move in the
Y-axis direction (front and back).
(U-SIC4R2/SIC4L2)
c) The tension of the X-axis and
Y-axis knobs is too tight or too
loose.
(U-SVRM/SVLM)
Repair request
If you cannot improve the phenomena after taking the above remedies, please contact Olympus for assistance.
At that time, please tell them the following information as well.
· Product name and abbreviation (Example: stage with coaxial knobs on the bottom right U-SVRM)
· Product number
· Phenomena
Set the main switch to (ON) while pressing the LIM switch and
the SET switch, and wait 5 seconds while pressing the LIM
switch and the SET switch.
Cause
The stage is not secured correctly.
The Y-axis is locked.
The tension of the X-axis and Y-axis
adjustment knobs is too tight or too
loose.
Remedy
Remedy
Secure the stage firmly.
Unlock the Y-axis.
Adjust the X-axis and Y-axis
adjustment knobs to the appropriate
tightness.
BX53M
Page
16, 30
Page
97
21
21
85