6.2 Calibration setting

This mode is used to set calibration concentration and actions. The calibration setting involves cali-
bration concentration, zero calibration, calibra-
tion range and auto calibration component/range.
Select the "Calibration Parameters", the screen
appears as shown at right.

6.2.1 Setting of calibration concentration

It allows you to set concentrations of the stan-
dard gas (zero and span) of each Ch used for
calibration.
(1) Select < User mode >  < Calibration
parameters >  < Calibration value >.
"Calibration Value Settings" screen ap-
pears as shown at right.
(2) Select the Ch you want to change by
pressing the
ENT
key and cursor moves preceding the
value.
(3) Select the concentration item you want to
set by pressing the
key (movable within the selected Ch).
Then press the
value is highlighted.
or the
key. Press the
,
key or the
ENT
key, and the selected
28
ENT
(
)
ENT
(
)