4.1.1 Zero/span calibration

Perform zero and span calibrations in the menu mode.
Caution on operation
• If zero/span calibration is attempted during warm-up operation, WARM-UP is displayed on the
upper right-hand corner of the screen, and the calibration cannot be executed.
• If zero calibration is not to be performed, press the
CH1 span calibration screen appears.
• Zero/span calibration can be performed either in measurement or standby mode. However, the
output hold is not executed in standby mode irrespective of the setting (kept in OFF state).
(1) Zero calibration mode
1) Press the
measurement 2 screen to display menu mode
screen.
2) Move the cursor to <1. Zero/Span Calibration>
using the
3) On the <Gas Select> screen that appears, select
zero gas conditions using the
Setting contents
• The setting screen varies depending on
calibration gas conditions. Perform calibra-
tion according to the following descriptions.
When <Air> is selected:
Zero gas: Clean air
Span gas: Dry gas in cylinder
When <Dry N2> is selected
Zero gas: Dry N2 in cylinder
Span gas: Dry gas in cylinder
4) Press the
screen appears.
5) Select ON/OFF of zero calibration using the
and the
Setting contents
ON
: Executes zero calibration.
OFF : Does not execute zero calibration.
6) Press the

4. SETTING AND MODE

4.1 Menu mode

key either on measurement 1 or
key, and then press the
key.
key, and the <Zero Cal.> mode
keys.
key.
key on the gas selection screen, and the
key.
MODE
CH1 Span
Calibration
screen
- 22 -
Measurement 1 or
measurement 2 screen
MODE
ESC
ENT
ESC
,
ENT
ESC
,
ENT
ESC
,
,
INZ-TN2ZSVS-E