Advertisement

4
Preparing the Fraction Collector

Swage Fittings

Swage Fittings

1
Select a nut that is long enough for the fitting you'll be
using.
3
Carefully slide the ferrule components on after the nut
and then finger-tighten the assembly while ensuring
that the tubing is completely seated in the bottom of the
end fitting.
1290 Infinity II Preparative Open-Bed Fraction Collector User Manual
2
Slide the nut over the end of the tubing or capillary.
4
Use a column or injection valve to gently tighten the fit-
ting which forces the ferrule to seat onto the tubing or
capillary.
NOTE
Don't overtighten. Overtightening will shorten the lifetime of
the fitting.
66