Degasser: cannot read signal

Error ID: 2243
The pump board gets no or wrong pressure signals from the built-in degasser.
Probable cause
1
2

Degasser: limit not reached

Error ID: 2244
This error is thrown, if the degasser does not become ready after 8 min, i.e. is
higher than 180 mbar.
Probable cause
1
2
3
1260 Infinity QuatPump VL User Manual
Degasser board defect, missing or not
connected to the pump main board.
Degasser sensor defect or not connected to
degasser board
Liquid in degasser tubing.
Leak in degasser tubing or chamber.
Degasser vacuum pump defect.
Error Information
Module Error Messages
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
7