List Sweep Output - Keysight B2980B Series User Manual

Femto/picoammeter electrometer/high resistance meter
Table of Contents

Advertisement

Function Details
Voltage Source (B2985B/B2987B)

List Sweep Output

The list sweep function is effective for performing an arbitrary waveform output.
For example, the voltage source can apply the waveform as shown in
and the B2985B/B2987B can perform the measurement at each output value. The
source output and the measurement can be performed in the minimum interval
100 s.
The voltage source output and measurement timing is controlled by the trigger
system. And the interval can be set to a constant value if the trigger type is set to
TIMER.
To set the trigger system, see
interval is set by the Period parameter. The number of output values is set by the
Count parameter.
To set the voltage source output values, use the List Sweep dialog box. See
sweep setup" on page
Figure 5-5
the output waveform.
To set the voltage output range, use the Spot Source Range field, see
Setup Sub-panel" on page
Figure 5-5
218
154.
shows an example setup of the List Sweep dialog box with an image of
List Sweep Dialog Box
"Trigger Setup Sub-panel" on page
151.
Keysight B2980B User's Guide, Edition 1
Figure
5-5,
157. The trigger
"List
"Range

Hide quick links:

Advertisement

Chapters

Table of Contents
loading

This manual is also suitable for:

B2981bB2983bB2985bB2987b

Table of Contents