Output Connection Dialog Box - Keysight B2980B Series User Manual

Femto/picoammeter electrometer/high resistance meter
Table of Contents

Advertisement

Volts
Common
(B2985B/B2987B) Displays the following softkey for setting the source and
measurement operations commonly.
Wait

Output Connection Dialog Box

(B2985B/B2987B) This dialog box provides the following parameters for setting
the source operation and connection.
Low Terminal State
Output-Off State
Keysight B2980B User's Guide, Edition 1
The softkey displays the present
Manual/Auto status of this function.
Pressing this key switches the status.
Absolute
Absolute value of resistance measurement
result. ON or OFF.
ON sets the resistance measurement result
to absolute format.
OFF cancels the absolute format.
The softkey displays the present ON/OFF
status of this function. Pressing this key
switches the status.
(B2985B/B2987B) Displays the following softkey for setting the
voltage measurement operation.
Attn.
Attenuation level (coefficient) for the voltage
input. 1 or 100.
When performing voltage measurement
with the Keysight N1413A/N1414A adapter,
set this level to 100 in order to return actual
voltage value.
The softkey displays the present status of
this function. Pressing this key switches the
status.
Source and measurement wait time setup. See
Dialog Box" on page
Voltage Source Low terminal connection,
CCOM (circuit common) or FLOAT (floating)
Output-off state, HIGH Z (high impedance),
NORMAL (normal), or ZERO (0 V)
Front Panel Reference
175.
Config Key Group
"Wait Control
173