Square Waveform Output

The voltage source can also perform a square waveform output.
To set the square waveform output, see
(B2985B/B2987B)" on page
To set the voltage output range, use the Spot Source Range field, see
Setup Sub-panel" on page
Figure 5-6

Ranging Mode of Voltage Source

The following two modes are available only when the voltage source is used as a
sweep source. In other cases, only FIXED mode is available.

Output-Off Status

The output-off status is automatically the condition of the voltage source,
immediately after its output is turned off. The status must be specified before
enabling the voltage source output. The available conditions are shown in
5-2.
To set the output-off status, see
Keysight B2980B User's Guide, Edition 1
Setup Parameters for Square Waveform
FIXED (fixed range)
The channel uses the specified range only.
BEST (best, only for the sweep source channel)
The channel automatically uses the minimum range which covers the whole
sweep output.
Voltage Source (B2985B/B2987B)
"Source Function Sub-panel
153.
Figure 5-6
shows the setup parameters visually.
151.
"Output Connection Dialog Box" on page
Function Details
"Range
Table
173.