Config Key Group

Press the Config key to display the following softkeys for setting several functions
of the meter and the voltage source.
Source
(B2985B/B2987B) Displays the following softkeys for setting the voltage source
operation.
Connection
Sweep
Measure
Displays the following softkeys for setting the measurement operation.
Connection
Filter
Ranging
Coulomb
Keysight B2980B User's Guide, Edition 1
Source operation and connection setup. See
Connection Dialog Box" on page
Sweep source detail setup. See
page
174.
(B2985B/B2987B) Measurement operation and connection
setup. See
"Input Connection Dialog Box" on page
Digital filter setup for measurement. See
Box" on page
175.
Automatic ranging speed setup, Normal or Fast
Normal
Performs the basic auto-ranging operation.
The channel automatically sets the range
which provides the best resolution in
performing the measurement.
Fast
For performing high throughput
measurement, reduces observation time to
determine measurement range. This setup
can be used when the measurement signal
noise (superposed from AC power, etc.) is
low.
The softkey displays the present Normal/Fast status of this
function. Pressing this key switches the status.
(B2985B/B2987B) Displays the following softkeys for setting
the automatic discharge function of the charge measurement.
Front Panel Reference
Config Key Group
"Output
173.
"Sweep Dialog Box" on
174.
"Measure Filter Dialog
171