Trigger Output - Keysight B2980B Series User Manual

Femto/picoammeter electrometer/high resistance meter
Table of Contents

Advertisement

Function Details
Trigger System
Figure 5-11

Trigger Output

B2980 can output trigger at the timing of Trigger output shown in
The trigger output terminal can be selected from the following terminals.
The trigger output can be set by using
page
5-7.
240
Transient and Acquire Device Actions
Transient device action
- for voltage output -
Source delay
Acquire device action
- for measurement -
Measure delay
Internal bus, INT1, INT2
LAN port
Digital I/O pin, EXT1, EXT2, EXT3, EXT4, EXT5, EXT6, EXT7
Trigger Output connector, TOUT
183. The trigger output timing and the setup parameters are shown in
Step value
Trigger
Measure
Trigger
"Trigger Configuration Dialog Box" on
Keysight B2980B User's Guide, Edition 1
Source delay
Step value
Trigger
Measure delay
Measure
Trigger
Figure
5-10.
Table