Self-Calibration - Keysight B2980B Series User Manual

Femto/picoammeter electrometer/high resistance meter
Table of Contents

Advertisement

Installation
Maintenance
To perform
The following procedure shows how to perform the self-test.
self-test
1. If it is in the remote state, press the Cancel/Local key.
2. Press the On/Off switch and confirm that the switch turns off.
3. Disconnect test leads and cables from the measurement and output terminals.
4. Press the following function keys.
5. Press OK. This starts the self-test.

Self-calibration

Keysight B2980 provides the self-calibration function to maintain the
measurement performance. If the environmental temperature changes  3 C or
more, perform the self-calibration. This is effective for the accurate measurements
by minimizing the effect of thermal drift. The self-calibration must be performed
after warming-up of 60 minutes. Before performing the self-calibration, turn the
voltage source output off and disconnect test leads and cables from the terminals.
To perform
The following procedure shows how to perform the self-calibration.
self-calibration
1. If it is in the remote state, press the Cancel/Local key.
2. Press the On/Off switch and confirm that the switch turns off.
3. Disconnect test leads and cables from the measurement and output terminals.
4. Press the following function keys. Confirmation dialog box is opened.
5. Press OK. This starts the self-calibration.
66
If you feel that the instrument may be defective
For preventive maintenance
System Menu > System > Cal/Test > Self-Test
Confirmation dialog box is opened.
System Menu > System > Cal/Test > Self-Cal
Keysight B2980B User's Guide, Edition 1