Trace Buffer

The trace buffer collects the test result data until a full buffer is detected. The
maximum data size is 100,000 blocks. The data flow is shown in
data block may contain multiple data, such as current measurement data, charge
measurement data, voltage measurement data, resistance measurement data,
voltage source output data, calculation result data, limit test data, time data,
temperature data, humidity data, and status data. They are selected by using the
Format keys of the I/O key group. See
For setting the trace buffer, see
In
If data is stored in the trace buffer, its statistical data can be calculated.
Calculable statistical data is as follows.
Keysight B2980B User's Guide, Edition 1
Figure
5-9, the variables indicate the following data.
VOLT: Voltage measurement data (B2985B/B2987B)
CURR: Current measurement data
CHAR: Charge measurement data (B2985B/B2987B)
RES: Resistance measurement data (B2985B/B2987B)
TIME: Time data (timestamp of the measurement start trigger)
STAT: Status data or limit test status
SOUR: Voltage source output data (B2985B/B2987B)
CALC: Math (calculation) result data or limit test data (= raw data  offset data)
TEMP: Temperature data (B2985B/B2987B)
HUM: Humidity data (B2985B/B2987B)
MEAN: Mean value
SDEV: Standard deviation
MIN: Minimum value
MAX: Maximum value
PKPK: Peak to peak value
"I/O Key Group" on page
"Trace Buffer Setup Dialog Box" on page
Function Details

Trace Buffer

Figure
5-9. One
190.
181.
233