Output Function; Output Range; Ac/Dc Mode - NF DP Series Instruction Manual

Programmable ac power source
Table of Contents

Advertisement

10.1

Output Function

Output function
10.2

Output Range

Output range
10.3

AC/DC Mode

AC/DC mode
AC
ACDC
DC
DP Series/DP Series Type R
Continuous, Sequence, Simulation
100 V range, 200 V range
Single-phase and Multi-phase models
AC, ACDC, DC
The signal source and amplification section of this product are AC-coupled, and
the DC component is canceled.
Only the AC output setting of 40 Hz or higher is available.
The signal sources that can be combined are INT, VCA, SYNC, EXT, and ADD.
When the waveform superimposed with DC is amplified using EXT and ADD
(using external signal sources), it may not be an intended output because the DC
component is canceled. In this case, select the ACDC mode.
The signal source and amplification section of this product are DC-coupled, and
the DC component is also amplified.
The AC and DC output settings of 1 Hz or higher are available.
The signal sources that can be combined are INT, SYNC, EXT, and ADD.
In the polyphase model, polyphase system, and polyphase output of the
Multi-phase model, only the AC setting is available.
Fixed to this mode in the Simulation.
Select this mode when you want to amplify a signal including DC, to superimpose
DC (DC offset), or output a frequency of 40 Hz or lower. Also select this mode
when the DC component temporally occurs, for example, by sudden change of
voltage or phase.
The signal source and amplification section of this product are DC-coupled.
Only the DC setting is available.
The signal sources that can be combined are INT and VCA.
It is unavailable for the polyphase model, the polyphase system, and polyphase
output of the Multi-phase model.
All models
All models
Polyphase models and Polyphase
systems
AC, ACDC
Description
10.1 Output Function
269