10. Specifications

10.19 Simulation

This allows you to simulate power line abnormalities, such as blackout, voltage rise, voltage drop,
abrupt phase change, and abrupt frequency change. It can be used only for AC, sine wave, and
ACDC-INT. Note that it does not support IEC or other standards test. When performing the tests
specified by the standards, use appropriate peripheral equipment. In the polyphase model,
polyphase system, and polyphase output of the Multi-phase model, only the balanced mode is
available.
Number of memories
Number of steps
Setting range of Step Time
Parameter
Simulation control
290
5 (non-volatile)
6 (Initial, Normal 1, Trans 1, Abnormal, Trans 2, Normal 2)
0.0010 s to 999.9999 s (0 s is available only for the Transition Step)
Output range
(The above item is common within the Simulation)
AC voltage
Frequency
Waveform (sine wave only)
Start Phase (excluding the Transition Step)
Stop Phase (excluding the Transition Step)
Synchronous step output (2bit)
Trigger output
Repeat count (1 to 9999 times or infinite)
Initial
Normal1