2.5.1 Single-phase Model
Connection procedure
The figure in the following procedure is the figure when using the cable holder (option). For cable
holder installation, see 7.6.
1. Remove the resin-molded output terminal cover and the cable holder (option) of the output
terminal.
2. Connect the output terminal to the load with a cable. Be sure to tighten the screws firmly. If you
want to ground the output terminal at one point only, ground the Lo terminal. The Hi terminal
should not be grounded. If the load has a ground terminal, connect it to the chassis terminal of
the output terminal of this product.
CAUTION
The output terminal that can be grounded is the Lo terminals. Do not ground the Hi terminal.
DP Series/DP Series Type R
2.5 Output and System Cable Connection
23