10.18 Sequence

Effective only for AC-INT, ACDC-INT, and DC-INT.
Number of memories
Number of steps
Setting range of Step Time
Intra-Step behavior
Parameter
*62 *63 *64 *65
Sequence control
*62: The output of AC voltage, Frequency and DC voltage of step 0 can be changed on Sequence Edit
View during output ON.
*63: For DC-INT, the AC phase voltage, frequency, waveform, Start Phase, and Stop Phase cannot
be set.
*64: The DC voltage cannot be set in the polyphase model, polyphase system, and polyphase output
of the Multi-phase model.
*65: The Phase angle can be set only in the polyphase model, polyphase system, and polyphase
output of the Multi-phase model. The Start Phase and Stop Phase are specified for the L1 phase,
and the component of the Phase angle setting is added to them for the other phases.
DP Series/DP Series Type R
5 (non-volatile)
Maximum 255 (for each sequence)
0.0010 s to 999.9999 s
Constant, Keep, Linear Sweep
Output range
AC/DC mode (The signal source is fixed to INT)
(The above 2 items are common within one sequence)
AC phase voltage, frequency, waveform
DC voltage
Start Phase
Stop Phase
Phase angle
Step termination
Jump count (1 to 9999, or infinite)
Specification of the Jump-to step
Synchronous step output (2bit)
Specification of the branch step
Trigger output
Start
Stop
Hold
Resume
Branch 1, Branch 2
All models
10.18 Sequence
289