About Installation Environment; Table 2-2 Heat Value Of Dp Series - NF DP Series Instruction Manual

Programmable ac power source
Table of Contents

Advertisement

2.2

About Installation Environment

To use safely and maintain the reliability, take the followings into consideration:
 Install the product where the exhaust heat will not damage surroundings.
 This product performs forced air cooling using a fan. Keep the front and rear, on which the air
inlet and outlet are located, away from the wall at least 50 cm to secure air circulation.
 This product exhausts the heat from the air outlet on the rear. Do not place heat-sensitive
objects around the outlet.
 If you install the product in a closed narrow room, air conditioning equipment is required. The
heat value of DP series is shown in Table 2-2. The heat value of DP series Type R is shown in
Table 2-3.
Model
DP015S
DP030S
DP045S
DP060S
Single-phase
DP075S
DP090S
DP105S
DP120S
DP030D
DP060D
Single-phase
three-wire
DP090D
DP120D
DP045T
Three-phase
DP090T
DP045M
Multi-phase
DP090M
Note: Estimate at power input single-phase 200 V or three-phase three-wire 200 V or
three-phase four-wire 380 V (phase voltage 220 V), rated output.
DP Series/DP Series Type R

Table 2-2 Heat Value of DP series

Rated output
1.5 kVA
3 kVA
4.5 kVA
6 kVA
7.5 kVA
9 kVA
10.5 kVA
12 kVA
3 kVA
6 kVA
9 kVA
12 kVA
4.5 kVA
9 kVA
4.5 kVA
9 kVA

2.2 About Installation Environment

Heat value
0.5 kW
430 kcal/h
0.9 kW
780 kcal/h
1.4 kW
1210 kcal/h
1.8 kW
1550 kcal/h
2.3 kW
1980 kcal/h
2.7 kW
2320 kcal/h
3.2 kW
2760 kcal/h
3.6 kW
3100 kcal/h
0.9 kW
780 kcal/h
1.8 kW
1550 kcal/h
2.7 kW
2320 kcal/h
3.6 kW
3100 kcal/h
1.4 kW
1210 kcal/h
2.7 kW
2320 kcal/h
1.4 kW
1210 kcal/h
2.7 kW
2320 kcal/h