Top Cover Replacement - GE LOGIQ E9 Service Manual

Hide thumbs Also See for LOGIQ E9:
Table of Contents

Advertisement

GE
D
5573152-100, R
IRECTION
8-5-5

Top Cover replacement

Manpower /
Time
One person /
15 minutes
Table 8-11
1. Power down the system.
2. Disconnect the mains power cable from the wall outlet and all Probes and External I/O Cabling.
3. Remove Side Covers.
Preparation Links (if you need more information):
• 4-2-3 "Power shut down" on page 4-6.
• 8-5-3 "Side Covers replacement" on page 8-28.
8-5-5-1
Top Cover removal
1.. Remove the two screws that secure the Top
Cover.
Remove the Top Cover and gently pull back
and up.
Re-installed to the screws to avoid damage
to the Rear Handle.
. 2
EV
Table 8-10
Manpower / Time and Tools
Refer to:
8-2-5 "Tools needed for servicing the LOGIQ E9" on page
Preparations and Preparation Links
Preparations - you must perform the following steps
Table 8-12
Top Cover removal
(Top Cover Screw placement as seen from above.)
Steps
Chapter 8 Replacement procedures
LOGIQ E9 S
Tools
Corresponding Graphic
M
ERVICE
ANUAL
8-5.
8 - 35