GE LOGIQ E9 Service Manual page 232

Hide thumbs Also See for LOGIQ E9:
Table of Contents

Advertisement

GE
D
5573152-100, R
IRECTION
7-5-8
Diagnostics for Service Diagnostics
Front End (Card Rack) diagnostics listed in this section are applicable for GFI configurations. For MRX
configuration. Also for diagnostics for GFI Configurations running R3.x.x and later, and for MRX
Configurations, see:. Also for
7-5-8-1
Digital Receive
See:
7-5-3 "Diagnostics Window Overview" on page 7-16
features.
Symptoms:
Intermittent general problems (communication between DRX and GFI)
Image artifacts - pixilated/jagged edges
1.) Select the Digital Receive checkbox to run all the tests, or select the individual subtests to run only
the selected tests.
2.) Enter the desired Loop Count (numeric value of 4 digits or less).
3.) Click Execute to run the test(s).
7 - 30
. 2
EV
7-5-12-15 "Diagnostics Window Overview - R3.x.x and later" on page 7-95
Section 7-5 - Common Service Desktop
LOGIQ E9 S
for a description of the looping and list
M
ERVICE
ANUAL