GE LOGIQ E9 Service Manual page 727

Hide thumbs Also See for LOGIQ E9:
Table of Contents

Advertisement

GE
D
5573152-100, R
IRECTION
10-3-1
How often should care & maintenance tasks be performed? (cont'd)
Abbreviations used in the Customer Care Schedule
D = Daily
Table 10-2
Air Filter Grid
AC Mains Cable
Cables and Connectors
User Interface
LCD Monitor, Touch
Panel and Probe holder
Mechanical parts
Control Console
movement
Trackball Check
Trackball Cleaning
Disk Drives
(Data Backup)
Safe Probe Operation
Probe Air bubbles
Probe connectors
Console Leakage
Current Checks
Peripheral Leakage
Current Checks
. 2
EV
W = Weekly
Customer Care Schedule
Item
Service at Indicated Time
Remove the filter grid and clean the air filter.
Inspect AC Mains Cable
Check if all cables are fixed well seated at
the correct position and if there is no
mechanical damage visible.
Clean alphanumerical keyboard, Functional
keys, Digital potentiometers, TGC-Shift
potentiometers. (vacuum cleaner, lukewarm
soap water on a soft, damp cloth)
Clean LCD Monitor surface and Probe
holder with a fluid detergent in warm water
on a soft, damp cloth.
Clean and inspect the mechanical function
of wheels, casters, brakes and swivel locks
as well as side door, foot rest, front and rear
handle, and monitor holder.
Remove Dust and Coupling gel.
Check Translation/Rotation and Height
Adjustment (Elevation)
Check proper operation
(Cursor movement X, Y direction)
Remove trackball ring; open the trackball
housing and take out the trackball to clean it
with soft tissue and screwdriver shaft.
Test Image filing (Archive) Import and
Export data capability (DVD/CD Drive)
Clean probes and probe cables and check
acoustic lens housing (cracks) and probe
cables. In case of mechanical damage,
don't use them! Danger: Safety risk for
operator and patient.
To detect air bubbles in filling liquid, shake
the probe carefully and check abnormal
noise.
Remove dust/dirt of all probe connectors.
Clean with vacuum cleaner if dust is visible.
Chapter 10 Care & Maintenance
Table
10-2:
M = Monthly
A = Annually
D
•*
LOGIQ E9 S
W
M
A
Or at least bi-weekly
Mobile Unit Check weekly
also after corrective
maintenance
Be careful not to get the cloth
too wet so that moisture does
not enter the loudspeakers,
TGC-Slider, or other keys!
Be careful not to get the cloth
too wet so that moisture does
not enter the entire system.
Mobile Unit Check Daily
more frequently at Mobile Units
If failure occurs go to trackball
cleaning.
Please record it in the systems
setup maintenance report
* save the image filing data
• •*
weekly or at least monthly on
DVD/CD depending on the
number of examinations
* or before each use
Also after corrective
maintenance or as required by
your facilities QA program.
Also after corrective
maintenance or as required by
your facilities QA program.
M
ERVICE
ANUAL
Notes
10 - 5