Data sheet for enclosure/chassis leakage current
Table 10-13 on page 10-19 shows a typical format for recording the enclosure/chassis leakage current.
Measurements should be recorded from multiple locations for each set of test conditions.
The actual location of the test probe may vary by Ultrasound system.
NOTE:
Values in italics font are given as examples only. Record all data in the Electrical safety tests
log.
Table 10-13 Typical data format for recording enclosure/chassis leakage
Unit under test___________________________ Date of test:_____________________
System
Power
NOTE:
Values in italics font are given as examples only.
. 2
EV
Test Conditions
Grounding/PE
Off
Closed
Off
Open
On
Closed
On
Open
Chapter 10 Care & Maintenance
Measurement/Test Point Location
Rear
Lower
Panel
Frame
Connector
LOGIQ E9 S
Probe
Main
Handle
M
ERVICE
ANUAL
10 - 19