D
5573152-100, R
IRECTION
Table 8-79
2.
Disconnect the Power/USB Cable (P3), the
(cont.)
HDMI (P10) and the USB (P3).
3.
Remove the Ground Cable from the
Adapter and the Bulkhead, if not removed.
Discard (the new FRU will include the cable
and mounting hardware).
4.
Early R5 production and earlier:
Remove the Plate. The plate is captured by
the two retaining pins that protrude from the
Adapter.
NOTE: A pin or pick can be used to lift the
plate slightly to remove.
8 - 128
. 2
EV
V2 Monitor Arm Assembly Adapter removal
Steps
Section 8-6 - Replacing Top Console Parts
LOGIQ E9 S
Corresponding Graphic
OLED Monitor - R6 and later
Ground Cable installed (R5 shown)
M