8-4-7
Loading the Application Software Only (cont'd)
NOTE:
All systems must have at least the Basic Option string for the software to run. Any single Option String
may enable multiple options (i.e., Basic).
Demo Option strings turn on all options for a limited period of time. Do Not leave a demo Option string
in a unit when it is sold to a customer.
8-4-7-1
Verifying the software versions
1.) From the Touch Panel, select Utility -> System -> About.
2.) Verify the software version. Ensure that the software version shown on the screen matches the
8-4-7-2
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
If all are successful, include the following debrief script: LOGIQ E9 Service Manual, Direction
5573152-100, Rev. 2. Equipment passed all required checks and is ready for use.
See:
Section
4-2-2
4-2-3
4-3-13
4-3-5
4-3-7
4-3-17
4-3-10
4-3-17-6
8 - 24
. 2
EV
Figure 8-13 Enter Option String
disk label.
Table 8-2
Loading the Application Software only Functional Checks
Functional Check
Power ON/Boot Up
Power shut down
Probe/Connectors Checks
B-Mode Checks
System CFM and PWD Checks
Mechanical Function Checks
Basic Measurements
Software DVR (Option) Configuration
Functional Checks
(if present)
Section 8-4 - Loading / Reloading / Upgrading the Software
Debrief Script
LOGIQ E9 Service Manual, Direction
5573152-100, Rev. 2. Equipment passed all
required checks and is ready for use.