GE LOGIQ E9 Service Manual page 302

Hide thumbs Also See for LOGIQ E9:
Table of Contents

Advertisement

GE
D
5573152-100, R
IRECTION
8-4-6-6
Loading Software Patches
If a CD with software patches is included, load the CD, restart the LOGIQ E9 and follow the instructions.
8-4-6-7
Verifying the software versions
1.) From the Touch Panel, select Utility -> System -> ABOUT.
2.) Verify software and hardware versions as recorded in
versions" on page
8-4-6-8
Verifying the software option key settings
NOTE:
Software Option Dongles are software version dependant. If a new software version has been
installed, follow installation instructions for that software. Usually, new software option strings
are needed. When re-installing the same software, the Software Option Strings should be
reused.
1.) From the Touch Panel, select Utility -> Admin -> System Admin.
2.) Verify that the Software Option String(s) (alphanumeric strings) in the S/W Option Dongle section
is (are) the same as recorded in
3.) If needed, enter any new or missing option key string(s) in S/W Option Key field and select Add.
8-4-6-9
Verifying the TCP/IP settings
1.) From the Touch Panel, select Utilities -> Connectivity -> TCP/IP.
2.) Verify that all settings on the TCP/IP screen match those recorded in
from TCP/IP screen" on page
3.) Select "SAVE SETTINGS."
4.) Select OK to continue.
5.) Reboot system.
8-4-6-10
Verifying the AE Title and Port settings
1.) From the Touch Panel, select Utilities -> Connectivity -> Device.
2.) Verify that all settings on the Device settings screen match those recorded in
— Recording the AE Title and Port settings" on page
3.) Select "SAVE SETTINGS."
4.) Select OK to continue.
8 - 18
. 2
EV
4-25.
Section 8-4 - Loading / Reloading / Upgrading the Software
Table 4-14 "Record Software and Hardware
Table 4-10 "Software Option Keys" on page
4-16.
4-17.
LOGIQ E9 S
M
ERVICE
ANUAL
4-22.
Table 4-2 "Record settings
4-2-11-4 "Connectivity