7.5.3
Connecting External PET Cables & New CT-PET Cables to New A36-1,
A36-2, A37-1 Panels
• The external PET cables and the new CT-PET cables are connected to the new A36-1, A36-
• Power cables, connected to the A36-2 bracket, should be routed under the legs crossbeam.
• Other cables should be routed on the external side of the legs crossbeam.
Note:
Do not route any cables on the internal or upper side of the legs crossbeam. Connect the cables so
that their minimum bending radius is 12.5 mm.
• The grounding cable is connected to the new PET left leg.
• Connect "Terminator RHK", P/N 2295966, to GAN1 A37 J12.
6
2, and A37-1 brackets as shown in
A36-1 Bracket

Figure 7-19: External Cables Connections to A36-1 and A37-1 Brackets

Figure 7-20: Power Cable Connections to A36-2 Bracket

Chapter 7 - Gantry
D
ISCOVERY
Figure
7-19.
A37-1 Bracket
LS S
S
M
YSTEM
ERVICE
ANUAL
Page 239