4 Functional Requirements

4.1 Electrical Requirements

Innova System require two specific power lines, two three phase for C1 cabinet (see Illustration
AC Interconnect without Fluoro UPS in
Connect the main power demand to the X–Ray generator – Jedi (C1 cabinet).
X–Ray generator is feed by a 126 kVA three phase power line.
For Generator Power Supply Features, see ref 60601.1 IEC.
NOTE:
In CE configuration, C2 cabinet is fed by C1 cabinet.
Table 3-14: ELECTRICAL REQUIREMENTS: System line voltage without Fluoro UPS
Peak Max Power
input (kVA)
150
Table 3-15: ELECTRICAL REQUIREMENTS: System line voltage with Fluoro UPS UL
Peak Max Power
input (kVA)
150
Table 3-16: ELECTRICAL REQUIREMENTS: System line voltage with Fluoro UPS CE
Peak Max Power
input (kVA)
150
NOTE:
PDB (CE) maximum rating is equal to 57 kVA.
PDB (UL) maximum rating is equal to 124 kVA.
For C2 Cabinet there are two positions for the shunts:
Chapter 3 Physical Requirements of Room
Frequency (Hz) Mains Voltage (V) Max. Line Impe‐
50 or 60
380 to 480 (3 ph, ±
10%)
Frequency (Hz) Mains Voltage (V) Max. Line Impe‐
60
480 (3 ph + N,
±10% )
Frequency (Hz) Mains Voltage (V) Max. Line Impe‐
50 or 60
380/400/415 (3 ph
+ N, ±10% )
Line impedance should be compliant with IEC 601.2.7 Refer to the table
Impedance for feeder line between Generator cabinet and Hospital in
Section 4, Power and Grounding
When a Fluoro UPS is or will be installed, a Neutral line is mandatory. An IT or Delta
configuration power input requires the installation of an isolation transformer (see
transformer specification below)..
Chapter 6, Section 1, Power
Peak Current (A) Average Current
dance (ohms)
0.09 to 0.12
228 to 180 for C1
Peak Current (A) Average Current
dance (ohms)
0.09 to 0.12
180 for C1
Peak Current (A) Average Current
dance (ohms)
0.09 to 0.12
228 for C1
Requirements.
Direction 5160944-13-1EN, Revision 2
Distribution).
(A)
16 to 13
(A)
13
(A)
16
Mains Wire
4 wires (3 ph +
GND).
Mains Wire
5 wires (3 ph + N
+ GND)
Mains Wire
5 wires (3 ph + N
+ GND)
Max Line
Chapter 6,
79