7. Repair information and repair instructions

7.1 General

Only perform repairs to MODUL CapnoVol at an ESD workstation!
Follow the safety instructions in the instructions
for use for MODUL CapnoVol.
Any handling of this device assumes in-depth
knowledge and observance of the instructions for
use and of the servicing and repair
instructions.
Only perform repairs described in these
servicing and repair instructions. This is the
only way to guarantee that
MODUL CapnoVol works perfectly.

7.2 Open the device

Materials and tools required
ESD workstation
Phillips screwdriver, size 2
Pointed pliers
1. Place the device on a non-slip surface and
unscrew the 4 screws
housing
.
24
2. Carefully pull off top part of the housing
and swing it away.
3. Pull the insulating cover out of the inside of the
device.
Note
The insulating cover is glued. If the insulating
cover (WM 22387) is damaged when pulled
out, it must be replaced before the device is
closed.
22
Repair information and repair instructions
from bottom part of the
50
23
Always perform a test after any repair or
servicing work (see "4. Test the device" on
page 8).
If you replace components or individual parts,
use only genuine
WEINMANN replacement
.
parts
Note
The item numbers listed in the text which
follows are identical to the item numbers in the
replacement partslist on Page 35 and in the
overview on Page 4.
24
Insulating
23
cover
50
50
24