9. Repairs: Information and Instructions

9.1 General

An ESD workplace is essential for making repairs to
the ACCUVAC Rescue.
No work should be performed on the appliance
without a thorough knowledge of the Operating In-
structions and the Service and Repair Instructions,
which must always be complied with.
ACCUVAC Rescue is only intended for the pur-
pose described (see "3.1 Purpose" on page 6).

9.2 Opening the device

Tools and equipment required:

Phillips screwdriver size 2.

1. Switch off the ACCUVAC Rescue.
2. Disconnect the aspirator from the external
power supply.
3. Remove the collection canister
accessories.
4. Unscrew the holder
canister.
5. Open the case by unscrewing the 6 cross-
head screws
. When opening the case, be
46
careful not to damage the silicone sealing
cord.
6. Carefully pull the front and rear case elements
apart.
and any
28
for the collection
29
When replacing components or individual parts,
be sure to use only original Weinmann parts.
A function check (see "7.2 Performing the function
check" on page 10) must be performed after every
repair.
When ordering the rear part of the case, please
state the appliance type, appliance number and
year of manufacture.