7. Repair information and repair instructions

7.1 General

Please perform repairs on SOMNOcomfort 2/SOMNOcomfort 2e/SOMNObalance/SOMNObalance e
at an ESD workstation!
Refer to the safety instructions in the instructions
for use for SOMNOcomfort 2/
SOMNOcomfort 2e/SOMNObalance/
SOMNObalance e.
Any handling of this device requires in-depth
knowledge of and adherence to the
instructions for use and the servicing and
repair instructions.
Perform repairs described in this servicing and
repair instructions only. This is the only way to
ensure that SOMNOcomfort 2/
SOMNOcomfort 2e/SOMNObalance/
SOMNObalance e work properly.

7.2 Tools and facilities

In order to perform the repairs described in this chapter, you will need the following tools and facilities:
Torx Screwdriver T 20
Phillips screwdriver PZ 2
Ball-headed Allen wrench 3 mm
Knife with a flat smooth blade for removing the front panel foil
Scissors or scalpel for removing motor mounts
Nail scissors or ticket-punch to mark the servicing plaque
ESD Workstation
Make sure that your hands and your
workspace are clean when performing
repairs.
After
repair a servicing check must be
every
performed (see "5. Final test" on page 15).
When replacing components or single parts,
use genuine Weinmann replacement parts
only.
When ordering the unit's lower part of the
housing
you have to state unit type, built
48
and serial number.
Note:
The item numbers stated in the following text are
identical with the item numbers of the Replacement
parts list on page 33 and the overview on page 4.
Repair information and repair instructions
21