Repair Information And Repair Instructions; General; Tools And Facilities - Weinmann Homecare MEDUCORE Easy Service And Repair Instructions

Automatic external defibrillator. device nos. 1000 to 1340 from device no. 1341 with ilcor 2005
Table of Contents

Advertisement

8. Repair information and repair instructions

8.1 General

8.2 Tools and facilities

24
Repair information and repair instructions
Please perform repairs on MEDUCORE Easy only on
an ESD work station!
Refer to the safety instructions in the operating
manuals for MEDUCORE Easy.
Any handling of this device requires in-depth
knowledge of and adherence to the operating
manual and the maintenance and repair manual.
Perform only the repairs as described in this
maintenance and repair manual. This is the only
way to ensure that the MEDUCORE Easy
operates properly.
Make sure that your hands and your workspace
are clean when performing repairs.
After every repair a maintenance check must be
performed (see "4. Checking the device" on
page 9).
When replacing components or single parts, use
only genuine WEINMANN Parts.
When ordering the device's lower encasing
you have to state the device type, year of
manufacture and serial number.
Note:
The item numbers stated in the following text are identical
to the item numbers of the spare part list on page 34 and
the overview on page 4.
In order to perform the repairs described in this
chapter, you will need the following tools and
facilities:
Phillips screwdriver size: PZ 2
Nail scissors or ticket-punch to mark the
maintenance plaque
Pincers
High-voltage protected needle nose pliers
ESD Workstation
Discharging device WM 40009
or:
protected high-voltage workstation
25

Advertisement

Table of Contents
loading

This manual is also suitable for:

Wm 40000Wm 40005