Field or Button
Properties button
302-42828-01 - Dec 2012
Click to view or edit the constants describing the physical properties of the
adsorbent and adsorptive.
These options are disabled if Entered is selected in the
Interaction Parameter group box.
Adsorbent group box -
Contains the parameters for the sample. If using Computed for the
interaction parameter, all fields are enabled. If using Entered, only the
values in the Diameter and Diameter at zero energy text fields may be
edited.
• Description dropdown list - click to select the name of the adsorbent
used in the analysis.
• Diameter - enter the diameter of the sample (or adsorbent) atom.
• Diameter at zero energy - enter the diameter of an atom at zero
interaction energy: (2/5)
Adsorptive group box -
Contains the parameters for the adsorptives (provided with the software
and/or user-defined). If using Computed for the interaction parameter, all
fields are enabled. If using Entered, only the values in the Diameter and
Diameter at zero energy text fields may be edited.
• Mnemonic dropdown list - click to select the mnemonic of the
adsorptive gas in use.
• Diameter - enter the diameter of the gas phase (or adsorbent) atom.
Description (continued)
1/6
× diameter.