3 Operate the Smart VacPrep
10. Install the safety shield.
11. Press the Select button for the port to be degassed. A blue LED is lit.
12. Press the Start button on the Smart VacPrep panel to start the degas process for all ports that
have a blue LED lit.
13. When degassing is complete and the sample has cooled, the sample can be transferred to an ana-
lysis port.
L
Q
S
OAD
UICK
TART
QuickStart Degas Conditions can be copied into a sample file. Reference the Degas Conditions section
of the analyzer operator manual.
Degas Conditions Fields and Buttons Table
Field or Button
Close
Degas Conditions
Heating Phase
3 - 4
D
C
EGAS
ONDITIONS INTO A
Description
Closes the active window.
Use to browse for a .DEG file that contains degas condition parameters
to be used in the analysis.
This option is applicable when degassing with either a Smart VacPrep or
a SmartPrep.
Enter up to five stages of degas conditions.
Temperature. Degas temperature.
l
Temperature Ramp Rate. Rate at which the temperature is to
l
change.
Time. Amount of time to heat the sample.
l
Use to modify the table contents.
Insert. Inserts one row above the selected row.
l
Delete. Deletes the selected row.
l
Clear. Clears all table entries and displays only one default value.
l
Append. Inserts one row at the end of the table.
l
S
F
AMPLE
ILE
Smart VacPrep Operator Manual — 067-42800-01
Dec 2013