Calibrating the System

Zero Pressure

Use this option to evacuate the system and zero the transducers.
1.
Install a blank sample tube on each applicable port.
2.
Go to Unit [n] > Calibration > Zero Pressure.
3.
Ensure that all transducers are selected, then click Start. The window closes when the operation is
complete.

Match Transducers

Use this option to zero and match the selected transducers to the main manifold transducer.
A sample or Po tube must be installed on each selected port before clicking Start.
9-26
TriStar II 3020
302-42828-01 - Dec 2012