Appendix B
2480-
File [name] cannot be analyzed. It is currently being edited.
Cause:
Action:
2481-
Error accessing the sample information file [file name].
Cause:
Action:
2482-
File cannot be opened for writing. It is already in use.
Cause:
Action:
2483-
An analysis cannot be performed on [file name]. It is open for editing and con-
tains errors.
Cause:
Action:
The edit session for [file name] must be saved before the analysis. Save
2484-
changes and continue with the analysis? [Yes, No]
Cause:
Action:
2485-
The service test file has an invalid status and cannot be used for this analysis.
Cause:
Action:
B-6
An attempt was made to start an analysis using a file that is open for
editing.
Finish editing the file, save and close it then start the analysis.
An unexplained error prevented access to this file.
The hard disk drive may be corrupt. Run diagnositics.
An attempt was made to open a file currently being used.
Locate the application using the file (in the Micromeritics application,
use the Windows menu item to get a list of all windows, one of which
may contain this file).
An attempt was made to use a sample file containing errors that is
currently open.
Go to the window containing the file, correct the errors and save it.
An attempt was made to start an analysis using a file that contains
unsaved changes and is open for editing.
Select Yes to save the changes and proceed with the analysis. Select
No to cancel the analysis and continue editing the Sample
Information file.
The selected service test file has a status other than No Analysis.
Select a different service test file or create a new one and use Replace
All to copy parameters from the file originally selected.
TriStar II 3020
302-42828-01 - Dec 2013
Need help?
Do you have a question about the TriStar II 3020 and is the answer not in the manual?
Questions and answers