8 E
M
RROR
ESSAGES
Program error messages are listed numerically. If the Action response indicates to contact a
Micromeritics service representative, record the error message, then make backup copies of any files
involved in the operation.
4407
Error searching for installed Smart VacPreps. The registry could not be read..
Cause: The application was not installed properly.
Action: Reinstall the application. If the problem persists, contact your Micromeritics service rep-
resentative.
4408
The %s in %s already has Smart VacPrep S/N %d. The Smart VacPrep
must be removed from the %s before it can be added.
Cause: The Smart VacPrep was already installed for another application / unit.
Action: Remove the Smart VacPrep from the installed unit before adding it to the preferred unit.
4409
A free IP address on the same subnet as %s could not be found."
Cause: All IP addresses on the network for the Ethernet card specified during installation are in
use by other Micromeritics applications on this computer.
Action: Uninstall unused Micromeritics applications.
Action: Configure a different Ethernet card for use by the application using the application
installer.
4410
The .INI file could not be updated with configuration for Smart
VacPrep S/N %s."
Cause: The application .INI file is opened by another application and could not be updated.
Action: Close all open applications and add the Smart VacPrep again using the Smart VacPrep
menu.
4411
Error dosing.
Cause: The backfill timed out.
Action: Ensure there is gas available and the pressure regulator is set to the appropriate pressure.
Also ensure that the gas supply regulator shut-off valve is open.
Smart VacPrep Operator Manual — 067-42800-01
Dec 2013
8 Error Messages
8 - 1
Need help?
Do you have a question about the TriStar II 3020 and is the answer not in the manual?
Questions and answers