Appendix B
2456-
Insufficient file handles available. Application cannot continue.
Cause:
Action:
2458-
An instrument is performing a critical operation. Wait a few moments before
exiting the application.
Cause:
Action:
2459-
An instrument is busy. A delay in restarting this application could result in loss
of new data. Continue with program Exit? [Yes, No]
Cause:
Action:
2460-
Fatal Communications error on [Unit n - S/N: nnnn]
Cause:
Action:
B-4
More than 50 files are open at the same time.
Refer to an operating system manual and set the limit for open files to
50 or greater.
An attempt was made to exit the application while the instrument was
performing a critical operation. This operation must be completed
before the application can be stopped.
Wait a short time and attempt to stop the application again.
An attempt was made to exit the application while an analysis was in
progress. While this is possible, the data collected when the
application is inactive will not be permanently recorded until the
application is re-started. A power failure to the instrument could
cause some data to be lost.
If not concerned with the potential for loss of data should a power
failure occur, click Yes to continue; otherwise click No.
There was a fatal error in communication between the application and
the software in the instrument. All displays for that instrument will be
closed.
Ensure that the analyzer is connected to the computer on the
communications port configured in the Setup program. Stop and
restart the analyzer software. Contact a Micromeritics service
representative if this error message continues.
TriStar II 3020
302-42828-01 - Dec 2013