TriStar II 3020
6021-
Servo calibration failed.
Cause A:
Action A:
Cause B:
Action B:
6022-
This file already selected for the analysis.
Cause:
Action:
6023-
The instrument [number] is not calibrated.
Cause:
Action A:
Action B:
6024-
Evacuation failed.
Cause:
Action A:
Action B:
302-42828-01 - Dec 2013
The maximum time was exceeded before the target pressure point
was reached. The nitrogen regulator may be set too low or turned off.
Set the analysis gas regulator to at least 10 psig (0.7 bar), then resume
the analysis.
The analysis gas bottle is empty.
Connect a new analysis gas bottle, then resume the analysis.
An attempt was made to choose a file for the analysis on this port
which has been selected for another port.
Choose a different file.
The analyzer application is in the process of initializing the
instrument and is unable to locate the calibration files.
Click OK. Select Unit [n] > Calibration > Load from File and select
a file containing calibration data.
Click OK. Close the application and use the Setup program to
reinstall calibration files.
While attempting to zero the pressure transducers, the instrument was
unable to evacuate to a pressure of less than 1 mmHg. This may be
due to a leak or a bad calibration.
Check the sample tube fitting and ensure that it is securely attached to
the port.
Use the Setup program to reinstall the calibration files.
Appendix B
B-31