Appendix B
4048-
Warning: An error occurred while restoring the heat of adsorption report editor.
Cause:
Action:
4049-
The sample [file name] does not have enough data. A minimum of two adsorp-
tion points is required.
Cause:
Action:
4050-
None of the requested quantities adsorbed is within the range of the collected
data of more than one sample file.
Cause:
Action:
4051-
The sample [file name] does not have any data in the range of the requested
quantities adsorbed.
Cause:
Action:
4052-
Fewer than two points are selected for this report.
Cause:
Action:
4053-
At least two data points must be selected for t-Plot calculations.
Cause:
Action:
B-22
The state of the heat of adsorption report editor could not be restored.
Default settings will be used.
No action.
A sample file has been included in the Heat of Adsorption report that
does not have enough data.
Remove the file from the selected file list.
The Heat of Adsorption report failed because the specified quantities
are not within the range of the collected data.
Edit the quantities adsorbed so that they are within the range of the
collected data or select other sample files.
The sample's data cannot be interpolated to any of the quantities
adsorbed.
Edit the quantities adsorbed so that they are within the range of the
collected data.
At least two points are required for the BET calculations.
Edit the calculation assignments for the BET report.
At least two points are required for the t-Plot calculations.
Edit the calculation assignments for the t-Plot report.