TriStar II 3020
Field or Button
Use isothermal
jacket checkbox
Use filler rod
checkbox
Vacuum seal type
option
Load from Sample
File button
Close button
Preview button
Save button
Degas Conditions File
File > Open > Degas Conditions (or click the Degas Conditions tab when in Advanced format)
The Degas Conditions tab provides details for setting up the Degas Conditions parameter file. This
information will be automatically applied during the degassing procedure if using the SmartPrep
Degasser. Use this option when the SmartPrep Degasser is installed.
Field or Button
Description text box
302-42828-01 - Dec 2012
Select if an isothermal jacket is to be used. An isothermal jacket maintains
a constant temperature profile along the sample tube stem during an
extended analysis of more than 1 or 2 hours.
Select if a filler rod is to be used in the sample tube. A filler rod reduces
the stem free space volume, resulting in reduction of free space error.
Select the seal type to be used.
If using calculated free space, enter the warm free space, cold free space
and non-ideality factor; or click Load from Sample File to import the
information from an existing sample information file.
Refer to
Common Fields and Buttons - File Menu
Contains the description of the current file.
Description (continued)
Description
Degas Conditions File
Options, page 5-1.
5-11