Appendix B
4058-
At least two data points are needed for Temkin calculations.
Cause A:
Action A:
Cause B:
Action B:
4059-
Fewer than 2 points available for MP-Method calculations.
Cause:
Action:
4060-
Sample [file name] contains no data points.
Cause:
Action:
4061-
The t-curve must contain at least 2 points.
Cause:
Action:
4062-
Error during report preparation.
Cause:
Action:
B-24
Less than two data points have been selected for the Temkin report; at
least two are required.
Select Temkin points on the Collected Data window. If calculation
assignments are not being used, edit the Temkin Report options,
Absolute pressure range in the sample file.
At least one point with a negative pressure value was selected to
include in the Temkin report.
Negative pressure points void Temkin calculations. Open the sample
file and click the Collected Data tab. Locate the negative values in
the Temkin column and deselect it.
At least two points are required for the MP-Method calculations.
Edit the calculation assignments for the MP-Method report.
An attempt was made to save a sample without collected data as a t-
curve or alpha-S curve.
Repeat the Save As t-curve or Save As alpha-S operation after
opening a sample that has collected data.
At least two points are required in a thickness curve definition.
Edit the thickness curve.
An internal processing error has occurred.
Contact a Micromeritics service representative if this error message
continues.
TriStar II 3020
302-42828-01 - Dec 2013