Bjh Adsorption/Desorption Report Options - Micromeritics TriStar II 3020 Operator's Manual

Hide thumbs Also See for TriStar II 3020:
Table of Contents

Advertisement

Report Options Files

BJH Adsorption/Desorption Report Options

The BJH calculation determines the mesopore volume/area distribution which accounts for both the
change in adsorbate layer thickness and the liquid condensed in pore cores. Reports can be generated
from both adsorption and desorption data.
The fields for both report options are identical unless otherwise specified. In the Selected Reports list
box, highlight BJH Adsorption (or BJH Desorption), then click Edit.
An incomplete pore distribution may be generated if a thickness curve selection is not
a good match for the sample being analyzed.
Field or Button
Thickness Curve
group box
BJH Correction
group box
Minimum / Maximum
BJH width text boxes
5-44
Refer to
Options, page 5-38.
t-Plot Report
Select the type of correction to apply to calculations. The selected type
displays in the report header.
Standard - uses original BJH models.
Kruk-Jaroniec-Sayari - good for reference thickness curves.
Faas - good for statistical thickness curves.
Enter the minimum and maximum diameter (radius or width) of pores to
include in the BJH reports.
Does not display on the
Desorption window.
Description