Stratasys J4100 User Manual page 106

3d printing system
Table of Contents

Advertisement

Stratasys J4100 User Guide
If the UV level measured is not acceptable, the wizard calibrates the UV lamp by
adjusting the current supplied to it, and then tests it again.
When all UV tests have finished, the results are displayed, showing the condition
of the lamps after calibration (see Figure 4-114 below).
Figure 4-114 Results and condition of UV lamps after calibration
After examining the UV calibration results:
1. Click Next.
2. Remove the UV sensor and close the printer cover. After confirming the items
DOC-26000 Rev. C
listed in the wizard screen, click Next.
Figure 4-115 UV sensor removal
If the results are not acceptable for quality printing, this is indicated in the final
wizard screen.
4 Operating and Maintaining the Printer
4-81