Stratasys J4100 User Guide
2. Lead the cable out of the printer so that you can close the printer, and tape it
3. Close the printer, while checking that the sensor does not move out of
4. Set the UV meter to measure mJ/cm2.
5. Set the range on the meter to 2,000.
6. Connect the cable from the sensor to the UV meter.
7. Turn on the UV meter, and wait until "0 0 0.0" appears on the display.
8. In the wizard screen, confirm that the sensor is positioned correctly, and click
Continue with "UV Measurement " on the next page.
DOC-26000 Rev. C
down to make sure that it does not interfere with the moving print block.
Figure 4-111 Correct UV-sensor placement
position.
Next.
4 Operating and Maintaining the Printer
4-79