7-14 | 5830 Moisture Analyzer

Analog Interface PCB

The Analog Interface PCB is located on top of the MCU PCB at the front
of the electronic module (Figure 7-9).

Removing the analog interface PCB

Remove the four knurled screws that hold the board in place.
Unplug the connectors attached to the board (Figure 7-10).
Carefully, lift the board until it slides free of the connector.

Replacing the analog interface PCB

Position the new analog interface PCB on the connector and press
firmly to insert.
Replace the screws removed earlier. Do not overtighten.
Plug in the analog interface board connectors.

MCU PCB

The MCU PCB is located below the Analog Interface PCB (Figure 7-6).

Removing the MCU PCB

Remove the analog interface PCB as described in the previous section.
Using a 1/16" hex wrench, remove the four standoffs that hold the
MCU board in place. Unplug all connectors (Figure 7-11).

Replacing the MCU PCB

Install the new MCU PCB on the mounting plate using the four stand-
offs removed and a 1/16" hex wrench.
Remount the analog interface PCB as described in the previous section.