System Test

Figure 4-7.
Main menu for System
Test key.
Use the System Test Key to perform tests on various components of the
analyzer.
Config
System Test
Test Display
Tests the display by providing a "checker board" pattern on the display.
Test Keypad
Tests the keypad by echoing characters to the display.
Test Alarms
Tests the alarm contacts by toggling them open / close.
Test Analog Output
Tests the analog output by allowing the user to set outputs to 4, 12 or 20 mA.
Test Serial Communications
Performs a "loop back" test and reports the result to the display.
Test
Display
Cancel
Test
Press Key
Keypad
Open Contacts
Test Alarms
Close Contacts
4 mA Output
Test Analog
12 mA Output
Out
20 mA Output
Test Serial
Start
Com
Cancel
Start
|
User Interface
4-13